GE Healthcare · X-ray VascularAdvantxInnova50108

Bolus run Aborted Failure ID ADXR KO: Bolus Chasing 1.0 Focal Spot Comment: Xrays Aborted after xx Images by e…

Replace C-Arc SMA board

Symptom

Bolus run Aborted Failure ID ADXR KO: Bolus Chasing 1.0 Focal Spot Comment: Xrays Aborted after xx Images by error 50211 Bolus run Aborted Failure ID ADXR KO: Bolus Chasing 1.0 Focal Spot Comment: Xrays Aborted after xx Images by error 50211

Error message

CARC error Code = 50108, error Text = POS SMA: power voltage failure, debug Text += pos-error=4AB9, hw-status=0002, recovery Class = DO, major Function /= LC Gantry C_ARC Degraded Motion displayed. Gantry Motion may or may not be inhibited. Error code 50108 HW = 0002 FRT_ARC. System impact: Degraded gantry motion. Displayed message : None. 0000 FRT_L 0001 FRT_PVT 0002 FRT_ARC 0003 FRT_II

Applies to

Models
Advantx, Innova
Product
Advantx E Monoplane Advantx E Systems advantx e, vamp Advantx E generator Systems