Bolus run Aborted Failure ID ADXR KO: Bolus Chasing 1.0 Focal Spot Comment: Xrays Aborted after xx Images by e…
Replace C-Arc SMA board
Symptom
Bolus run Aborted Failure ID ADXR KO: Bolus Chasing 1.0 Focal Spot Comment: Xrays Aborted after xx Images by error 50211 Bolus run Aborted Failure ID ADXR KO: Bolus Chasing 1.0 Focal Spot Comment: Xrays Aborted after xx Images by error 50211
Error message
CARC error Code = 50108, error Text = POS SMA: power voltage failure, debug Text += pos-error=4AB9, hw-status=0002, recovery Class = DO, major Function /= LC Gantry C_ARC Degraded Motion displayed. Gantry Motion may or may not be inhibited. Error code 50108 HW = 0002 FRT_ARC. System impact: Degraded gantry motion. Displayed message : None. 0000 FRT_L 0001 FRT_PVT 0002 FRT_ARC 0003 FRT_II
Applies to
- Models
- Advantx, Innova
- Product
- Advantx E Monoplane Advantx E Systems advantx e, vamp Advantx E generator Systems
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.