GE Healthcare · MRExcite225721022527202252721
SRF and TRF Power Faults Due to MGD Power Supplies.
Symptom
Scan Stops in Middle of Scans See Long Description Hard failure during scanning, APS works ok, AP recon error watching APS window.
Error message
2257210 SCP did not respond to the start packet of WavRec.This may be due to a TPS Reset failure 2257210 SCP did not respond to the start packet of WavRec. This may be due to a TPS Reset failure 2252720 Sequence playout error. The SRF signaled a SRF power fault error. see long description Error: 2252720 Host: lx-mr Process: NSP File: AGP:agp drv.c Line: 510 Sequence playout error. The SRF signaled a SRF power fault error. 2252721 - The SRF signaled a TRF power fault error. Error: 2252721 Host: lx-mr Process: NSP File: AGP:agp drv.c Line: 519 Sequence playout error. The SRF signaled a TRF power fault error.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2294300-8 MGD Power Supply ***OBSOLETE PART Case # 11430551*** See Long Description ***OBSOLETE PART Case # 11430551*** This part has been identified as technically obsolete and no short term solution has been identified. FE must identify and order a higher or lower level spare as required. Contact your support engineering, On-line Engineering or Parts ID team should you require support. 212012173 7-24-254 Suitable alternative part is 5181495-3. Please see latest appropriate FRU manual for product in question. 212002950 February 20, 2009.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.