GE Healthcare · X-ray VascularInnova3004830050
Xray not available; Reset system to recover, Call Service is displayed in the Image monitor.
Symptom
Xray not available; Reset system to recover, Call Service is displayed in the Image monitor.
Error message
30048, error Text = POWER UNIT : KVmax FAILURE, debug Text = HWStatus=0034 User=0000 0000 42E4, recovery Class = DO, major Function /= ECM System impact: None. 1 tube spit that can be heard by customer. Displayed message : None. ADTX error 42E4. 30050, error Text = POWER UNIT : THYRISTOR TQ FAILURE, debug Text = HWStatus=0036 User=0000 0000 42E6, recovery Class = DO, major Function /= ECM System impact: None. 1 tube spit that can be heard by customer. Displayed message : None. ADTX error 42E6.
Applies to
- Models
- Innova
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems
- Software
- $, Innova 2000 Service Pack FMI F4, HerSvcPack_2.02 HerSvcPack_2.02 DL Applications version 11.23.26 Ghost DL-1 "F" : 17.4 Ghost DL-1 "G" : 15.5 Ghost DL-2 : 16.4 Ghost DL-4 : NA DL : 11.23.26 DL Atlas Core : 3.1O-1 RTAC Atlas Core : 3.1O-2 IDC : 2.13 Calypso : 11.13_FMI_7 Calypso Uptime : 11.25_FMI Titan : H0.57 Pos. Motorola : 5.2Y3-2 Pos. Roadrunner : 5.0R
- Component
- vamp inverter 600 volt ac supply MPPU1 A4 ac supply panel 2115403 vamp inverter 600 volt ac supply MPPU1 A4 ac supply panel 2115403 vamp inverter 600 volt ac supply MPPU1 A4 ac supply panel 2115403
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.