GE Healthcare · CT101821444510414449200
Lightspeed 1.X - Octane
Symptom
The image works browser would lock up intermittently and some images transferred to pacs would be corrupted.
Error message
Error: 10182 File IO error during file read for save read error See the LD for complete error Error: 10182 Host: fbct Process: arsjob File: ArchiveDevice.c Line: 7391 File IO error during file read for save : read error, errno=5, /export/home1/sdc_image_pool/images/p43/e43/s91/i2995.CTDC.4 2 In SYSLOG on the OC: Dec 17 15:44:27 6A:fbct unix: dksc0d2: <6>[Alert] Media error<6>: Unrecovered data block read error (asc=0x11, asq=0x0)<6>, (data byte 36)<6>, Block #14445104<6> (14449200)<6> Dec 17 15:44:27 6A:fbct unix: dksc0d2: <6> retrying request
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2169940-12 9Gb disk drive 2169940-12 9Gb disk drive Part #: 2169940-72 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 1.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 9GB 10K RPM DISK WITH SLED FRU: Yes Status: Orderable 2169940-12 9Gb disk drive Device 0 2 Disk SGI IBM DNES-309170Y FW Rev: SA30 Part #: 2169940-72 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 1.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 9GB 10K RPM DISK WITH SLED FRU: Yes Status: Orderable Comments: FLS - OCTANE 2 9GB DISK DRIVE USED ON HELIOS 2 AND 3 AND LX 9/25/03 CHECK WITH FIELD TO SEE IF THEY CAN USE THE 18GB PART# 2169940-78 IF NONE OF THE 9GB ARE AVAIL/WE HAVE BEEN ABLE TO GET THE 9GB FROM CPU/ MS10665 GELS - SEE FLS COMMENT:2169940-12 MAY BE USED INSTEAD (RK/PMT 5/2003) THE WRONG FIRMWARE IS LOADED ON THE DRIVES FLORIANE 09/09/03 ALL STOCK MUST BE PURGE AND SEND IT BACK TO COMPUTER PARTS UNLIMIFLORIANE09/09/03,OCTANE 2 9GB DISK DRIVE
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required