GE Healthcare · CT101821444510414449200200000424200000416200000415
Lightspeed 1.X - Octane
Symptom
Images fail to display and recon shutdown. Images fail to display and recon shutdown
Error message
Error: 10182 File IO error during file read for save read error See the LD for complete error Error: 10182 Host: fbct Process: arsjob File: ArchiveDevice.c Line: 7391 File IO error during file read for save : read error, errno=5, /export/home1/sdc_image_pool/images/p43/e43/s91/i2995.CTDC.4 2 In SYSLOG on the OC: Dec 17 15:44:27 6A:fbct unix: dksc0d2: <6>[Alert] Media error<6>: Unrecovered data block read error (asc=0x11, asq=0x0)<6>, (data byte 36)<6>, Block #14445104<6> (14449200)<6> Dec 17 15:44:27 6A:fbct unix: dksc0d2: <6> retrying request Error: 200000424 Image available space failed. See the long description for the full error Sat Jan 22 18:25:36 2005 Error: 200000424 Host: chr09bc1cti Process: dms File: AvailSpace.cxx Line: 155 Image available space failed. code: 200000416 Or Sat Oct 28 06:55:39 2006 Error: 200000424 Host: ct02 Process: dms File: AvailSpace.cxx Line: 155 Image available space failed. Error code: 200000415 Error: 200000437 Image installation failed for image (for some Exam/Series/Image number) Open Long Description for full error varaiations Sat Jul 16 13:11:02 2005 Error: 200000437 Host: hsa2 Process: dms File: InstallImage.cxx Line: 316 Image installation
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required