Scanner locks up intermittently

Symptom

The image works browser would lock up intermittently and some images transferred to pacs would be corrupted.

Error message

Error: 200002745 Stat failure after image closed See LD for details Tue Jun 7 12:27:18 2005 Error: 200002745 hostid: su101 process: imagecreate ic_imagecreate_req.c 536 Stat failure after image closed Image Path : /usr/g/sdc_image_pool2/reserved/IMSse802zHG/E302U0_S2U0_I7U0.11 18165232_474766 File stat status : -1 Errno : 5

Error: 10182 File IO error during file read for save read error See the LD for complete error Error: 10182 Host: fbct Process: arsjob File: ArchiveDevice.c Line: 7391 File IO error during file read for save : read error, errno=5, /export/home1/sdc_image_pool/images/p43/e43/s91/i2995.CTDC.4 2 In SYSLOG on the OC: Dec 17 15:44:27 6A:fbct unix: dksc0d2: <6>[Alert] Media error<6>: Unrecovered data block read error (asc=0x11, asq=0x0)<6>, (data byte 36)<6>, Block #14445104<6> (14449200)<6> Dec 17 15:44:27 6A:fbct unix: dksc0d2: <6> retrying request

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2368697 Linux DDR Workstation - Used on GOC2 only See long description for details Part #: 2368697 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: LINUX DDR WORKSTATION - GEMS CT USA FRU: Yes Status: Orderable Comments: FLS 3/5/05 (212001261 Shawn Fox) HOST COMPUTER (HP). THIS IS THE GOC2 HP 8000 COMPUTER . ***EXTREMELY IMPORTANT: NEW SOFTWARE, S/W LOAD PROCEDURE AND S/W OPTION KEYS ARE REQUIRED IF THE HP8000 COMPUTER IS REPLACED. Contact the OLC to assist with new Option Keys.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.