GE Healthcare · MRExcite225437550164
Found open center conductor in one of the two ground isolation boxes.
Symptom
Unable to scan with body or Surface Coils with RF Errors No receive signal when scanning body or surface coils. Head scans ok.
Error message
2254375 RF Amplifier Reflected Power Fault. Sun Oct 12 17:12:24 2008 Error: 2254375 Host: MR02_OC02 Process: NSP File: SCP:Srfd3RfNbCanHWO Line: 2289 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: Spin Echo Coil: BODY Scan: 50164/1 RF Amplifier Reflected Power Fault. RF Amplifier reflected power exceeds specification. Possible causes: disconnected or defective output cable from RF Amplifier, or defective RF Amplifier.->
Applies to
- Models
- Excite
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 14.0 For Hino Excite HDe 14.0 systems only.
- Component
- 5118952 1.5T Ground Isolation Filter See Long Description Description in FEMC: Ground Isolation Filter These are the Isolation Boxes that connect to the BODY QUADRATURE HYBRID ASSEMBLY
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.