O2 Alarm Sensor is Defective.
Symptom
Oxygen Alarm Sounds Intermittently and Can Be Reset The Oxygen (O2) alarm sounds intermittently and the operator is able to reset it. Sometime later it will go off again. Oxygen Alarm Sounds Intermittently and Can Be Reset. The Oxygen (O2) alarm sounds intermittently and the operator is able to reset it. Sometime later it will go off again.
Applies to
- Models
- Excite, Signa, Discovery, Optima
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2112207 ISA 40M OXYGEN SENSOR 2112207 Oxygen monitor sensor FEMC Comments: FLS - REPLACES 46-305001P2. FE WILLNEED TO ORDER 46-320936P1 RETROFIT PLATE KIT IF NOT PREVIOUSLY INSTALLED IN ORDER TO USE THIS NEW OXYGEN SENSOR PER JOHN FERUT, 1-95 IF ORDERING FOR MOBILES THAT HAVE THE OLD STYLE METAL RF CEILING BOX# 46-306602G1, IT SHOULD BE REPLACED WITH NEW CEILING BOX# 2124496 (WHICH ALREADY HAS THIS SENSOR INSIDE). HOWEVER, IF THE MOBILE HAS A REMOTE SENSOR MODULE ON THE CEILING (46-307676P2), THEN THE NEW CEILING BOX IS NOT NEEDED. *MUST ALWAYS HAVE AVAILABLE AS THIS IS A SAFETY ISSUE PART ! JOE BANGO, JR, AT SUPPLIER KNOWS ABOUT THE ADJUSTMENTS REQUIRED AT THE SITES... JL 7/96 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.