GE Healthcare · MR2240010

The SRP was not issuing the TTL logic control signal to the Preamp which enables the RX Coil Protection Circuit.

Symptom

Auto Prescan Failure

Error message

2240010 Coil switch bias board error

Applies to

Product
Profile 3 Ultrasparc CPU with AN8290Y gradients
Software
Profile Contour 7.67 Applies to Profile1 through 4 and Contour1 through 4. This is the last revision of software for this product line. Requires 7.66 LFC. 7.67 is a patch.
Component
2161974 or 5114871-2 SRP Assy 2161974 upgraded to 5114871-2. Same functionality as an SRI. See SN DOC0625410 Rev.1 dated July 7, 2009.