Pular para o conteúdo principal
GE Healthcare · MRExcite22627252224666

Profile 5 shoulder coil failing autoprescan with open circuit bias faults.

Symptom

Profile 5 PA Shoulder Array coil failing autoprescan with open circuit bias faults

Error message

2262725 Phased Array bias error(open) is occurrred 2262725 Host: profile Process:NSP File: SCPCoilIDSrpHWO Line:538 Phased Array bias error(open) is occurrred

2224666 Auto Prescan failed, Pre amplifier is not ready 2224666 Host: profile Process: psc File: PscMainControl.cc Line: 1311 Auto Prescan failed, Pre amplifier is not ready

Applies to

Models
Excite
Product
Profile with Excite
Software
Profile5 Applies to Profile5 (Excite) only. Profile5 is 12.x based software, not ProfileHD.
Component
2261603 P9364 VM Base Plate FLS- baseplate for shoulder array and YMS Profile Wrist array coil. Can be ordered from MRI Devices VC 479600 MS08992
Area
MR