Symptom

OrthoMRI Error Starting DSP code See Long Description Notice: Always use the most current revision of service documentation. DSPM: performs the digital generation of the System control signals, perform the image reconstruction applying the FFT on the received raw data as well, implements the parallel digital 3M connection for camera printers. Technical data: – Average power consumption1.5 A @ 3.3 V, 0.05 A @ 5 V – Two micro processors – PCI bus – DSP0 (21161@100 MHz), 400 MFLOPS, 100 MIPS, RAM 128 KB internal cache, 0.39 MB – DSP1 (21161@100 MHz), 400 MFLOPS, 100 MIPS, RAM 128 KB internal cache, 256 MB or 512 MB DRAM

Error message

OrthoMRI Error starting DSP Code 43 Try to perform these steps in this order: re-boot the system and check the DSPM initialization, change DSPM card, reload the software See Long description for additional information. Always use the most current revision of service documentation. Try to perform these steps in this order: re-boot the system and check the DSPM initialization, DSPM utilization via block diagram illustration.

Applies to

Product
OrthoMRI E-Scan (Windows) NE
Component
OrthoMRI DSPM board LNR9102498000
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.