Pular para o conteúdo principal
GE Healthcare · MR

Elevated vessel pressure due to degraded cold head performance. hfoi2 HePressure Limit.

Symptom

hfoi2 alarm I2 HePressure Limit. hfoi2 alarm I2 HePressure Limit, the Helium vessel Pressure has exceeded the high alarm limit setting on the Magnet Monitor 3.

Error message

hfoi2 alarm I2 HePressure Limit. hfoi2 alarm I2 HePressure Limit, the Helium vessel Pressure has exceeded the high alarm limit setting on the Magnet Monitor 3.

Applies to

Product
Magnet HFO. Software Version Concept should NOT be selected for Magnet Products Serial # (Txxx) Field strength: 0.7 T He Vessel Pressure >0.9 psig but <6.25 psig Shield Temp for Cryo-cooler coldhead < 40 K Temp for Shield-cooler Coldhead < 35 K Cryocooler 2nd stage Temp > 3.8K but < 4.5K Recondenser Temp > 4.2K but < 4.6K Delta Temp (Coldhead - Recondensor) < 0 but < 0.3K Min. LHe Fill Level 75% of total vessel Min. LHe Ramp Level 100% Helium Volume 340 Liters Burst Disk 2253802-2 Baffle Assy 2226764
Component
SecondOpinion Part Number 2194003-2 Sumitomo Cold Head and Gasket Kit. See comments in resolution concept for second opinion requirements before ordering this part. See comments in resolution concept for second opinion requirements before ordering this part. Second Opinion Part Number 2244334 SINGLE STAGE CONFIGURED COLD HEAD. Second Opinion Part Number 2244334 COLDHEAD, SINGLE STAGE CONFIGURED. Used exclusively on HFO (OpenSpeed) 0.7T Magnets. See comments in resolution concept for second opinion requirements before ordering this part. Second Opinion Part Number 2244334 COLDHEAD, SINGLE STAGE CONFIGURED. Used exclusively on HFO (OpenSpeed) 0.7T Magnets. See comments in resolution concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues