Coil ID fails due to defective Quick Disconnect box with Coil ID.

Symptom

Quad Head Coil is not recognized by System See Long Description No green ID light, customer is unable to scan with this coil.

Error message

2218630 The download of a prescribed series failed. 2218630 Host: Andys3T Process: scn File: RxManagerIF.m Line: 556 The download of a prescribed series failed. Internal status code is 2218501.

2219712 The executable download of psd failed. The Host PSD and Hardware PSD are not compatible Wed May 14 20:36:06 2008 2219712 1 lx-mr scn RxMgrDownload.m 1288 The executable download of psd failed. The Host PSD and Hardware PSD are not compatible

2217483 The Host PSD and Hardware PSD are not compatible Wed May 14 20:36:06 2008 2217483 2 lx-mr scn PsdError.m 324 The Host PSD and Hardware PSD are not compatible

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2384268 CLASSIC HEAD COIL CBL ASM W/COIL ID See Long Description and Attachment See Attachment for the engineering drawing of this part: 2384268ADW Rev. 2. This is not available online. To confirm this is the most appropriate drawing, go to ePDM.
Area
MR