Coil ID Fails on All Multi Coils Due to 8 Channel Switch.

Symptom

Coil ID Fails on All Multi coils

Error message

2262134 SRI coil validation error. Coil ID is required for this connector. See Long Description Thu Jun 12 17:16:22 2008 Error: 2262134 Host: GOC-3TMR Process: NSP File: SCP:SRI Line: 326 SRI coil validation error: Coil ID device was not detected on a coil. Coil ID is required for this connector. Please try connecting coil again. Connector: Head/Multicoil Port->

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
5137103-2 8 Channel Crosspoint Switch Coil ID Signa HDe 1.5T Service Methods Rear Pedestal 5159685 Signa HDe 1.5T Service Methods Rear Pedestal 5159685 See YMS/Beijing MR Service Documen t website for replacement procedures.
Area
MR

Full cause & step-by-step resolution