GE Healthcare · X-ray VascularInnova37020383702011
Xray not Available - Reset System; If Problem Persists Call Service is displayed on the Image Monitor.
Symptom
Xray not Available - Reset System; If Problem Persists Call Service is displayed on the Image Monitor. Xray not Available - Reset System; If Problem Persists Call Service is displayed on the Monitor.
Error message
INX1X1: 3702038 Lat: Jedi: kvctl configuration problem on boot of the system 3702038 IH XRTechMgr Lat: Jedi: kvctl configuration problem Debug: jedi-error-code=705h JEDI Error Code: 0705H Configuration error Data Related to Error: 0 Tube 1 70C: OK, Tube 40C: OK, Rotor Thermal Error: OK, Inverter thermal Error: OK, Tank thermal Error: OK, Door Status: Closed Phase: STAND_BY This error will usually be followed by 3702011 IH XRTechMgr Lat: Generator configuration error on boot of the system 3702038 IH XRTechMgr Lat: Jedi: kvctl configuration problem Debug: jedi-error-code=705h JEDI Error Code: 0705H Configuration error Data Related to Error: 0 Tube 1 70C: OK, Tube 40C: OK, Rotor Thermal Error: OK, Inverter thermal Error: OK, Tank thermal Error: OK, Door Status: Closed Phase: STAND_BY This error will usually be followed by 3702011 IH XRTechMgr Lat: Generator configuration error on boot of the system
Applies to
- Models
- Innova
- Product
- Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.