GE Healthcare · MRExciteSigna14149
Unable to Print Due to Ethernet Hub in the GOC.
Symptom
Cannot DICOM Print to Kodak Camera. If the Profile console is connected directly to the camera, it will print. If the console is connected to the hospital network, images will not print. Ping reports the camera is alive.
Error message
14149 Failed to Print on DICOM Printer Tue Aug 27 07:15:31 2002 Error: 14149 Host: lx-mr Process: prsserver File: prsjob.c Line: 783 Failed to Print on DICOM Printer
Applies to
- Models
- Excite, Signa
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 5139480-2 8 PORT GIGABIT ETHERNET SWITCHING HUB See long description for important note FEMC Description: 8-Port Switch Kit with 4 cushions To view the most recent appropriate and detailed documentation, click to go to SERVICE METHODS. Note: BE CAREFUL when ordering this part. Per the FRU manuals, some products use this part for GOC and/or System Cabinet and some don't. Some products are supposed to use the 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT as the FRU. Please check the FRU manual for your system to see whether this part should really be used on the Product and for the purpose (i.e. GOC and/or Sys Cab).
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.