Scan Abort occurs occasionally, must press Move to Scan button on the Keyboard to continue.

Symptom

Scan Pause occurs occasionally, must press Move to Scan button on the Keyboard to continue May occur when the cradle has to reposition itself axially during the scan per prescription.

Error message

-06- Error Displayed on the Magnet Display. Cradle accuracy slippage This error is not recorded in the system log, only displayed on the Magnet Display.

Applies to

Product
Profile 4
Software
7.8x Profile4 or ProfileHSR only
Component
U0120AP Cradle Encoder Axis Jointer Coupler See Resolution concept for replacement procedure
Area
MR

Full cause & step-by-step resolution