Unable to Scan Using Invivo 7-Channel Breast Coil due to Customer using 2_BREAST Coil Name instead of 7breast both MRI.
Symptom
Unable to Scan Using Invivo 7-Channel Breast Coil Customer unable to scan using the Invivo 7-Channel Breast Coil. When they try this they receive the error message shown in the Long Description. Note that Invivo's Operator manual for this coil used to direct the customer to use a modified "2_BREAST" coil config for this coil. A newer revision of their manual has different instructions. Also please note that GE does not service this coil as it is sold directly by Invivo to the customer. Customer unable to scan using the Invivo 7-Channel Breast Coil. When they try this they receive the error message shown below. Note that Invivo's operator manual for this coil used to direct the customer to use a modified "2_BREAST" coil config for this coil. A newer revision of their manual has different instructions. Also please note that GE does not service this coil as it is sold directly by Invivo to the customer. Error: 2220007 Host: mr02 Process: scn File: RxMgrDownload.m Line: 2143 The prescribed RF coil and detected RF coil do not match. Upper coil connect status:0, Upper coil code:Null, Lower coil connect status:1, Lower coil cod e:WFdtKBdrLUmJKY398qJocJG3XcRNtyYd, Rx coil name:2_BREAST, Rx coil code:p1vf g3d42e3afd74fnok9d3cc310526v
Error message
2220007 The prescribed RF coil and detected RF coil do not match. See long description on reading/interpreting this type of error message Note: An example of this error message, with explanation, is provided below. Wed Mar 3 08:18:58 2004 Error: 2220007 Host: lx-mr3 Process: scn File: RxMgrDownload.m Line: 2074 The prescribed RF coil and detected RF coil do not match. 040303104912225 Upper coil connect status:0, Upper coil code:Null, Lower coil connect status:1, Lower coil code:q8dCUW0PD2EaETIplAp90PLrACM6CW38, Rx coil name:MRI 8HRKNRR PA, Rx coil code:q8dCUWOPD2EaETIpIAp9oPLrACM6CW38 NOTE: Upper coil connector status refers to connector plugged into upper port at "dog house" (i.e. head coil or Med' Adv' QUAD KNEE ("chimney") coil). Lower coil connect status refers to all PA coils / any coil using the lower port at the "dog house". Coil code should match that of what was detected (in this case the zeros were config'd as O and o (oh's - upper and lower case), and the number one following "ETIp" was an "I").Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.