Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova180001218000161800018

Database Inconsistency Failure

Symptom

Database Inconsistency Failure: Continue and call service for repair is displayed on the image monitor. This issue may occur during "Start" of an exam or during an exam. It occures when a file corruption has occurred in the DL database. This is a warning to the customer that file corruption has occurredon the image disk and a corrupt sequence may not network or display properly. Xray is not inhibited. Database Inconsistency Failure: Continue and call service for repair is displayed on the image monitor. This issue may occur during "Start" of an exam or during an exam. It occures when a file corruption has occurred in the DL database. Database inconsistency failure and a Debug text of Check And Repair Requested at Browser initialization. When the customer boots the system, the image monitor displays Database inconsistency failure: Continue and call service for repair. This is a warning to the customer that file corruption has occurredon the image disk and a corrupt sequence may not network or display properly. Xray is not inhibited.

Error message

IN2K: 1800012 ERR C80 PROC AR XRExpSens DSP errorTask C80 ERROR LEVEL 0 - USER ERROR: HEROIC ERROR NO ENDEC SYNCHRO 1800012 ERR C80 PROC AR XRExpSens DSP errorTask Component(s) RTAC C80 Boards 1 0 Async error - Card 1 (IP2) - C80 ERROR LEVEL 0 - USER ERROR: HEROIC ERROR NO ENDEC SYNCHRO - param:-1 c80error.cpp 473

Innova: 1800016 ERR C80 MEZZANINE CRITICAL Async error - Card 1 (IP2) - C80 ERROR LEVEL 1 - Mezzanine: critical receive error 1800016 ERR C80 MEZZANINE CRITICAL DO XRExpSens DSP errorTask Component(s) RTAC C80 Boards 1 0 Async error - Card 1 (IP2) - C80 ERROR LEVEL 1 - Mezzanine: critical receive error ( param 640 ) c80error.cpp

IN2K: 1800018 ERR SIB IP BOARDS LINKS IP1 (card 0) -IP2 (card 1) link failure detected 1800018 ERR SIB IP BOARDS LINKS IH XRExpSens DSP init Component(s) RTAC SIB C80 Boards links 1 0 [DSP automaticNaming] IP1 (card 0) -IP2 (card 1) link failure detected dsp driver.cpp 635

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues