GE Healthcare · X-ray VascularInnova10501421050145
No gantry motion is reported by the customer.
Symptom
No gantry motion is reported by the customer. The Gantry does not drive. Emergency backout works
Error message
INX100: 1050142 Anti-Cinematic Error 4B10 HW 0000 USER = 4000 - This error is due to big variation detected on Positioner Position values. It can be due to a shortcut on cable, or a bad potentiometer or encoder. 1050142 Anti-Cinematic Error 4B10 HW 0000 USER = 4000 - This error is due to big variation detected on Positioner Position values. It can be due to a shortcut on cable, or a bad potentiometer or encoder. HW code 0x0000 = Larm Axis User code 4000 = Position Problem INX100: Lift MCB PRD Code ACH - EXT Brkp Brake voltage Failure (PRD: AC) Lift MCB PRD Code ACH - EXT Brkp Brake voltage Failure. All Other MCB's held in PRD fault - LARM = BD; Pivot = BD, Arc = AD and Lift = AC; Troubleshooting - Troubleshoot the Axis with the Lowest PRD failure. INX100: 1050145 Anti-Cinematic Error 4B10 HW 0003 User = 8000 -This error is due to big variation detected on Positioner Position values. It can be due to a shortcut on cable, or a bad potentiometer or encoder. 1050145 Anti-Cinematic Error 4B10 HW 0003 User = 8000 -This error is due to big variation detected on Positioner Position values. It can be due to a shortcut on cable, or a bad potentiometer or encoder. HW 0003 = Lift Axis User 8
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.