GE Healthcare · MRExciteSigna22488212248850
Defective Analogic RF amplifier.
Symptom
Auto Prescan failed, RF Amplifier is not ready Getting another message that RF Amplifier is powering but message never goes away.
Error message
2248821 Analogic RF Amplifier 1 Device Fault 44. RF FET device is shorted. Replace amplifier Mon Nov 27 10:03:08 2006 Error: 2248821 Host: VDMRI Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2218 Analogic RF Amplifier 1 Device Fault 44. RF FET device is shorted. Replace amplifier.-> 2248850 RF Amplifier fault 71 Mon Nov 27 10:10:04 2006 Error: 2248850 Host: VDMRI Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2218 RF Amplifier fault 71.
Applies to
- Models
- Excite, Signa
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- 2224562-2 1.5T RF AMP POWER MODULE W/ FRU PKG (SRFD)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.