Pular para o conteúdo principal
GE Healthcare · MRExciteSigna2257359

Unable to Start Propeller with Error Code 2257359 Logged due to First Octet of TPS/MGD Subnet IP Address Same as Host PC…

Unable to Start Propeller with Error Code 2257359 Logged due to First Octet of TPS/MGD Subnet IP Address Same as Host PC's IP Address.

Symptom

Unable to Start Propeller Scan with Error Code 2257359 Logged See Long Description iTrak # 13202476 After the FE finished installing FMI 60746 "Upgrade from Signa HD_M5 to Signa HD_M5A" the customer reported she was unable to start Propeller scans. The [Start Scan] button would illuminate on the SCIM but nothing happened when they pushed it. The button would stay lit and the error code 2257359 shown below would be logged. This error was logged about two or three times every time they tried pushing the [Start Scan] button. Finally the techs would give up and hit the [Stop Scan] button just to get out of the situation and go on to something else. This would result in the error 2255008 "Auto Prescan failed, Stop button was pressed" being logged but this isn't really part of the problem, it's just a consequence of what the tech had to do to get out of this state. Thu Aug 7 15:23:31 2008 2257359 1 GEHS LxMxConsumer ScanControlCmdProcessor.cpp 129 Server Name: SC SCSM Scan command ignored because it arrived before previous stop. Thu Aug 7 15:25:24 2008 2255008 1 GEHS LxMxConsumer gemri.mx.prescancontrol.gemri.mx.prescanc 0 Auto Prescan failed, Stop button was pressed.

Error message

2257359 SCSM Scan command ignored because it arrived before previous stop. See Long Description 2257359 1 GEHS LxMxConsumer ScanControlCmdProcessor.cpp 129 Server Name: SC SCSM Scan command ignored because it arrived before previous stop.

Applies to

Models
Excite, Signa
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues