GE Healthcare · X-ray VascularInnova360208538003313602016

Ingrid errors 3800331, 3602085, 3602087, May see a message, Warning.

Symptom

Ingrid errors 3800331, 3602085, 3602087, May see a message, Warning. Run time 21sec. May see message like; Acquisition in 4 sec. Release switch to abort. 3800331 The innova system reports bad actual techniques during an aquisition, 3602085 Max spit error KV regulation code 8. 3602087 7 kV reg on kV rise are all regulation related. These were non tube related spits and replacing the Ingrid fixed the problem. May see a message, Warning. Run time 21sec. Acquisition in 4 sec. Release switch to abort.

Error message

KV REGULATION SPITS. 3602085, error Text = Jedi: too many tube spits, debug Text += jedi-error=316h, Spit Max error, data related to error: 8, recovery Class = AR, major Function /= XRTechMgr KV REGULATION SPITS. System impact: X-ray unavailable. Displayed message : "Xray aborted - Restart Xray; if problem persists, call Service". May be preceded by error 3800331 Bad actual techniques recieved in aquisition, and followed by 3602016 Degraded Image Quality

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
Innova Jedi Innova Jedi Innova Jedi
Area
X-ray Vascular