MRIQ Bright White Line in middle of Image and Gradient Hammer Test Failures Due to STIF Board.
Symptom
MRIQ Collapsed Image Bright White Line Down or Across Middle of The Image
Error message
2253850 Gradient Link Hammer test failed. The GP received an invalid X axis data value via the gradient data link. See long description for entire error. Sun Jun 19 00:22:58 2005 Error: 2253850 Host: lx-mr Process: diagCoord File: grad_hammer.c Line: 286 Diagnostic: Grad Hammer Iteration: 8 of infinite Action: 1 Task ID: 0x3d5c6c8 Time: 2541257 Object: SCP Test Task Gradient Link Hammer test failed. The GP received an invalid X axis data value via the gradient data link. Data Exp = 0007ffffH or 00080000H Data Rd = 00080010H Check the gradient data and clock fiber optic cable connections
Applies to
- Models
- Excite
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 14.0 For Hino Excite HDe 14.0 systems only.
- Component
- 2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.