Pular para o conteúdo principal
GE Healthcare · MRExcite22538502541257

MRIQ Bright White Line in middle of Image and Gradient Hammer Test Failures Due to STIF Board.

Symptom

MRIQ Collapsed Image Bright White Line Down or Across Middle of The Image

Error message

2253850 Gradient Link Hammer test failed. The GP received an invalid X axis data value via the gradient data link. See long description for entire error. Sun Jun 19 00:22:58 2005 Error: 2253850 Host: lx-mr Process: diagCoord File: grad_hammer.c Line: 286 Diagnostic: Grad Hammer Iteration: 8 of infinite Action: 1 Task ID: 0x3d5c6c8 Time: 2541257 Object: SCP Test Task Gradient Link Hammer test failed. The GP received an invalid X axis data value via the gradient data link. Data Exp = 0007ffffH or 00080000H Data Rd = 00080010H Check the gradient data and clock fiber optic cable connections

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
Area
MR

Full cause & step-by-step resolution