GE Healthcare · X-ray VascularInnova5011850211
The system will not complete a bolus chase.
Symptom
The system will not complete a bolus chase. Tube collision sensor error is displayed on the the monitor.
Error message
50118 POS Table: Encoder, potentiometer failure; Debug: , HDW STATUS=0000 or HW 0001,USR STATUS=XXXX followed by 50211 POS Table: Motion control error during Bolus procedure. Both occur at the same time during a Bolus run. 50118 POS Table: Encoder, potentiometer failure; Debug: POS ERROR=4AC3, HDW STATUS=0000,USR STATUS=XXXX 50211 POS Table: Motion control error during Bolus procedure. Debug: [XRImChainPos::logPosError] Reporing the Motion Control error for GantryMotionControlError See LD for complete error. 50118 POS Table: Encoder, potentiometer failure; Debug: POS ERROR=4AC3,HDW STATUS=0000,USR STATUS=XXXX. HW and User status indicate Longitudinal Feedback pot problem. - Longitudinal Feedback pot problem 50118 POS Table: Encoder, potentiometer failure; Debug: , HDW STATUS=0001,USR STATUS=XXXX followed by 50211 POS Table: Motion control error during Bolus procedure. Both occur at the same time during a Bolus run. This indicates a problem with the Lateral Table Feedback Pot During a bolus run. Recalibrate the Table Lateral Potentiometer P403 Check the alignment and correct any mounting hardware issues 50211 POS Table: Motion control error during Bolus procedure. Debug: [XRImChainP
Applies to
- Models
- Innova
- Product
- X Innova 4100 X Do not use---------4100 systems 4100 systems
- Software
- Bolus M4 DL release : DL_LBYBOLUSM4_9.2.050615 Ghost DL-2 : NA Ghost DL-3 : 18.8 Ghost DL-5 : 20.6 DL : DL_LBYBOLUSM4_9.2.050615 DL Atlas Core : 3.1.R RTAC Atlas Core : 3.1.R IDC : 4.5.5 Calypso : 2.21 Calypso Uptime : BOLUS1.24 Positioner : 1.2.24B_IS6.6 Jedi : B6.30BFC
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.