Rocker Tool not functioning correctly.

Symptom

Rocker Tool not functioning correctly. BPM's decrease after download when using Rocker Tool. See Long Description Have found recently while using 12.0 s/w that the Rocker Tool doesn't appear to be functioning correctly. When I use my usual technique that generally produces 13-15 BPM, it comes up first at a rate of 22-26 BPM. Not until I download the scan, does the BPM drop down into the 15 BPM range. The issue is that the respiratory BPM changes when I download the prescription. The prescription is a resp-gated FSE scan with the resp trigger set at 40%. It turns out the trigger level of 40% is what is causing the resp BPM rate changes. The waveform looks very regular, the period never changes. Before the download, I was seeing a lot of double triggers, one right after the other. These were raising the BPM rate. When the level went to 40% I didn't see any double triggers. I think there were still some occasionally because we saw the BPM spike on some triggers. They weren't noticeable on the display though. MRIhc00522

Applies to

Models
Excite
Product
3T Excite HD 12.0 software only
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR