New 3T with Excite installation.

Symptom

Excite 3T System Fails To Reset TPS During Installation

Error message

2219803 Could not register the object "theFreewheel" with Mr Mail. 1094564014 2 1 Tue Sep 7 13:33:34 2004 2219803 4 CHMMR scn SrxFreewheel.m 144 Could not register the object "theFreewheel" with Mr Mail.

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

200002356 dsserver died unexpectedly, no action being taken. Mon Nov 22 10:46:20 2004 Error: 200002356 Host: MRI1-OC0 Process: hostMonitor File: ssProcStop.c Line: 906 dsserver died unexpectedly, no action being taken.

2220013 Failed to Receive Coil Info Pkt from SCP Mon Nov 22 10:58:23 2004 Error: 2220013 Host: MRI1-OC0 Process: scn File: SrxCoilID.m Line: 124 Failed to Receive Coil Info Pkt from SCP

2257210 SCP did not respond to the start packet of WavRec. Mon Nov 22 10:57:41 2004 Error: 2257210 Host: MRI1-OC0 Process: WavRec File: WavRec.cc Line: 175 SCP did not respond to the start packet of WavRec. This may be due to a TPS Reset

2249424 Communication problem with the host. Simple Network Time Protocol (SNTP) request failed. see long description 1094564028 0 1 Tue Sep 7 13:33:48 2004 2249424

Applies to

Models
Excite
Product
3T Excite G3 Software ONLY
Software
G3 For 3T Excite only
Area
MR

Full cause & step-by-step resolution