GE Healthcare · MRExcite2244577
Current distortion fault due to SGA Assembly.
Symptom
System pauses with Current Distortion Fault Intermittently scans will pause with Z axis Current Distortion Fault, sometimes scan can be resumed and sometimes TPS reset gets scan going again.
Error message
2244577 The GP reported a % Axis Current Distortion Fault 2244577 "The GP reported a Z Axis Current Distortion Fault. If no other faults occurred, check LCoil Calibration. Else, run Gradient Driver Tests to isolate the FRU. If Tests pass, see Troubleshooting Manual."
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2250375-4 CONTROL_BOARD ASSEMBLY_SGA 2250375-4 CONTROL_BOARD ASSEMBLY_SGA 2250375-4 Part Desc: CONTROL_BOARD ASSEMBLY_SGA
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.