2244653 The GSC indicated a CPU reset during normal operation Due to the Gradient Switch Module.
Symptom
MRIQ Gradient Switch Related Errors Pause Scans and Resultant Images Contain a Bright Zipper or Band Artifact Scans suddenly pause then the system lights the Start Scan button for the customer to restart the scan. If this is done the images show a big zipper-like artifact that looks like a partially collapsed gradient. This is seen on many different scans not unique to one PSD. The problem may occur several times a day, and it may skip some days altogether, then reappear again another day. See the Attachment for copies of the errors and an example image. Scans suddenly pause then the system lights the Start Scan button for the customer to restart the scan. If this is done the images show a big zipper-like artifact that looks like a partially collapsed gradient. This is seen on many different scans not unique to one PSD. The problem may occur several times a day, and it may skip some days altogether, then reappear again another day. See the Attachment for copies of the errors and an example image. This system is a 1.5T Excite II TwinSpeed running 11.0 M4 software with Service Pack 1 loaded. Scans suddenly pause then the system lights the Start Scan button for the customer to restart the scan. If this is done the images show a big zipper-like artifact that looks like a partially collapsed gradient. This is seen on many different scans not unique to one PSD. The problem may occur several times a day, and it may skip some days altogether, then reappear again another day. There are several error messages logged which seem to point to the GSC (Gradient Switch Controller). Below are copied the error messages logged at the time of one of these scan pauses and a typical image. The fix was to replace the "TWIN GRAD SWITCH BOX" part number 2253466 in the TAC cabinet. This module contains the GSC. Reference SPR # MRIhc16978. Mon Sep 25 10:12:27 2006 2244648 WHMR1 NSP SCP:GP 668 The GSC reported a problem with the Whole switch on the Y axis. If the problem persists, please call a service engineer. Mon Sep 25 10:12:28 2006 2244655 WHMR1 NSP SCP:GP 1307 The GP received an ALARM message from the GSC. Subsequent error messages should reveal the cause of the alarm. Mon Sep 25 10:12:35 2006 2244598 WHMR1 NSP SCP:GP 1069 The gradient system on the Y axis recovered successfully. There are no faults at the present time. Mon Sep 25 10:12:35 2006 2244598 WHMR1 NSP SCP:GP 1069 The gradient system on the Z axis recovered successfully. There are no faults at the present time. Mon Sep 25 10:12:35 2006 2244598 2 WHMR1 NSP SCP:GP 1069 The gradient system on the X axis recovered successfully. There are no faults at the present time. Author Paul Lyman SME John Schumacher
Error message
2248108 APS received data over range error from DRF, on chan= n. 2248108 TOiMR2 NSP APS:dataMoverTask 604 Exception Class: Idle Severity: Unknown Function: Acquisition/Pulse Generation: Acquisition PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 APS received data over range error from DRF, on chan = 2 Need to reduce receiver gain. 2244653 The GSC indicated a CPU reset during normal operation. TPS reset may be necessary. See Long Description Error: 2244653 Host: GE3TMR Process: NSP File: SCP:Grad Line: 683 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: FRFSE-XL Coil: HDCTL 456 Scan: 190/3 The GSC indicated a CPU reset during normal operation. TPS reset may be necessary. If this error is persistent, check the GSC for loose or dirty internal power connections. Check communication connections.
Applies to
- Models
- Excite, Signa
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- 2253466 TWIN GRADIENT SWITCH Module Box Same part number for 1.5T and 3.0T. Same part number for 1.5T and 3.0T.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.