Unable to Scan with 3T MKS RF Amplifier Due to RF Amplifier.
Symptom
Unable to Boot or Power Up 3T MKS 35 kW RF Amp. Amplifier's LCD Display is all Black Squares. RF Amp Serial Link Errors The customer hit the Emergency Stop button which powers off the RF Amplifier. Afterward, the RF Amp could not be booted up. It didn't power up successfully. The RF Amp display is all “little black squares”, or all backlit characters. The FE checked all the fuses, including the three "hidden" fuses (See Service Note 67103 "35kW Amp Fuse Replacement" for details on these fuses), and none were blown. See Attachment for additional information. The customer hit the Emergency Stop button which powers off the RF Amplifier. Afterward, the RF Amp could not be booted up. It didn't power up successfully. The RF Amp display is all “little black squares”, or all backlit characters. The FE checked all the fuses, including the three "hidden" fuses (See Service Note 67103 "35kW Amp Fuse Replacement" for details on these fuses), and none were blown. See Attachment for additional information.
Error message
2261230 The SCP received an unknown frequency identifier from the RF cabinet. Please perform a TPS Reset. Frequency = 0xFF-> 2261230 PAC3T NSP SCP:RfNbCanHWO 1737 The SCP received an unknown frequency identifier from the RF cabinet. Please perform a TPS Reset. Frequency = 0xFF-> 2254342 RF Amplifier Interface RF Amp serial link failure. RF AIF failed to receive a packet transmission echo. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 5040 Expired time: 40 Tx byte: M 2254342 PAC3T NSP SCP:../rf_nb_serial 1795 RF Amplifier Interface <-> RF Amp serial link failure. RF AIF failed to receive a packet transmission echo. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 5040 Expired time: 40 Tx byte: M 2254347 RF Amplifier Interface RF Amp serial link failure. RF AIF failed to receive a packet transmission echo on 2nd try. If link is broke, a separate error message will be logged. If not, a spurious error occurred. Timer: 5080 Expired time: 40 Tx by 2254347 PAC3T NSP SCP:../rf_nb_serial 1947 RF Amplifier Interface RF Amp serial link failure. RF AIF failed to receive a pack
Applies to
- Models
- Excite
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion MKS 3T 35kW RF Amp 5104896 See comments in resolution concept for second opinion requirements before ordering this part.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.