Pular para o conteúdo principal
GE Healthcare · MRExcite22480163275422245894062796

1.5T Extremity Coil (Quad Knee/Foot Coil w/o cable) was causing error 2224589 and grainy images.

Symptom

MRIQ Grainy Images on Extremity Coil

Error message

2248016 Warning MR Signal is too high (overrange) Reduce R1 DRF ID 1 FE ID 0 channel 0 2248016 Host: mr waltham Process: NSP File: APS:DRF Line: 563 Warning: MR Signal is too high (overrange); Reduce R1 DRF ID: 1 FE ID: 0 channel: 0 arg1: 32754

2224589 TG reached the high limit. See Long Description Suite: MR01 Host: OC0 Proc: MSG_HANDLR0 Error: 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan. Note: On later systems, this same error code has been slightly reworded to read as follows: Thu Jun 12 11:06:12 2008 Error: 2224589 Host: MENEMR Process: NSP File: TPS:aps_exec Line: 4062796 TG reached the high limit. TG has been set to 190. Check patient positioning and if desired try Auto Prescan again or use Manual Prescan.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2225478-8 1.5T EXTREMITY COIL (QUAD KNEE/FOOT W/O CABLE) See Long Description 1.5T EXTREMITY COIL (QUAD KNEE/FOOT W/O CABLE) CABLE ASM = 2225478-9 PHANTOM = 2225478-10 FUSE = 2225478-14 VELCRO STRAP KIT = 2225478-7
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.