GE Healthcare · MRExcite22480163275422245894062796
1.5T Extremity Coil (Quad Knee/Foot Coil w/o cable) was causing error 2224589 and grainy images.
Symptom
MRIQ Grainy Images on Extremity Coil
Error message
2248016 Warning MR Signal is too high (overrange) Reduce R1 DRF ID 1 FE ID 0 channel 0 2248016 Host: mr waltham Process: NSP File: APS:DRF Line: 563 Warning: MR Signal is too high (overrange); Reduce R1 DRF ID: 1 FE ID: 0 channel: 0 arg1: 32754 2224589 TG reached the high limit. See Long Description Suite: MR01 Host: OC0 Proc: MSG_HANDLR0 Error: 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan. Note: On later systems, this same error code has been slightly reworded to read as follows: Thu Jun 12 11:06:12 2008 Error: 2224589 Host: MENEMR Process: NSP File: TPS:aps_exec Line: 4062796 TG reached the high limit. TG has been set to 190. Check patient positioning and if desired try Auto Prescan again or use Manual Prescan.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2225478-8 1.5T EXTREMITY COIL (QUAD KNEE/FOOT W/O CABLE) See Long Description 1.5T EXTREMITY COIL (QUAD KNEE/FOOT W/O CABLE) CABLE ASM = 2225478-9 PHANTOM = 2225478-10 FUSE = 2225478-14 VELCRO STRAP KIT = 2225478-7
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.