GE Healthcare · MRExciteSigna2252302

2252302 Multicoil Bias Fault due to defective coil cable

Symptom

Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.

Error message

2252302 Multicoil Bias Fault Detected - Transmit Fault 2252302 IC_OPEN_SG NSP SCP:DCB 793 Multicoil Bias Fault Detected - Transmit Fault. Try re-seating all coil connectors. MCD1 Channel 5 of 16, counting from 1 Value: -2 mA Limit: 0 mA Mask of failed channels: 0x10 Ch 1-8 : Legacy MC Conn Ch 17-24 : Conn A Ch 25-32 : Conn B->

Applies to

Models
Excite, Signa
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
2417162 ASSY,FRU,CABLE,VALUE,BODY ARRAY (cable for coil M3335MC Signa 1.5T HD 8-Channel Body Array). See Long Description Note Even though the cable has "VALUE" in the part name, this is misleading because it isn't just used on HDe, it's also used on 1.5T HDxT, MR450, and MR450W as can be seen in the Coil Vendor Manuals sections of the respective Service Methods documentation as well as in the Coils Compatibility Matrix (Cable is a FRU for coil M3335MC Signa 1.5T HD 8-Channel Body Array). To view the most recent appropriate and detailed documentation, click to go to the Engineering Documents .
Area
MR