GE Healthcare · MRExcite2256460

SCP2 board not installed in MGD Chassis.

Symptom

Propeller Scan Option Will Not Run or Runs Very Slow propeller option will not function

Error message

2256460 Switch to Monarch failed due to wrong hardware configuration. SCP2 is required. See Long Description Error: 2256460 Host: LXMR Process: LxMxConsumer File: SCSwitcher.cpp Line: 101 Server Name: SC Switch to Monarch failed due to wrong hardware configuration. SCP2 is required.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2294300-12 SCP2 Board See Long Description 2294300-12 SCP2 Board. FEMC Parts Comments states as follows: If no stock then 2294300-19 is a suitable substitute. 212002950 February 8, 2010.
Area
MR

Full cause & step-by-step resolution