Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova2800027

The system is not sending sequences to the DICOM network destination

Symptom

The system is not sending sequences to the DICOM network destination The system is not automatically sending images to the network destination. The customer indicates that the images have not been sent to the remote destination. The system is not auto sending images to the network destination. The customer indicates that the images have not been sent to the remote destination.

Error message

Innova: mwlServer.log reports DJR - root error= 4112; final Error=1004 when attempting to establish connection with the broker Innova: mwlServer.log reports DJR - root error= 4112; final Error=1004 when attempting to establish connection with the broker. F0=DJR - root error= 4112; final Error=1004; Message= MC Open Associati on F0=Attempting to establish connection: AEtitle(xxxxxxx), IPaddr(xx.xx.xx.xx),

error Code = 2800027, error Text = Dicom push failed, debug Text += Association timeout, recovery Class = DO, major Function /= XRSubSysMgrDL System impact: Push failed. Displayed message : None.

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues