TG Reached the High Limit Due to the 3T RF Body Coil. 3T Excite HDx 14.0 M5.

Symptom

Auto Prescan Fails with TG Reached the High Limit See error in Long Description 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan.

Error message

2224713 No successful FTG check in X iterations APS1 will run automatically. Error: 2224713 Host: DMHMRI Process: NSP File: APS:aps_exec Line: 1 No successful FTG check in 2 iterations. APS1 will run automatically.

2224666 Auto Prescan failed, Please Try Again. Error: 2224666 Host: signa1 Process: psc File: PscMainControl.cc Line: 1351 Auto Prescan failed, Please Try Again.

2224589 TG reached the high limit. See Long Description Suite: MR01 Host: OC0 Proc: MSG_HANDLR0 Error: 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan. Note: On later systems, this same error code has been slightly reworded to read as follows: Thu Jun 12 11:06:12 2008 Error: 2224589 Host: MENEMR Process: NSP File: TPS:aps_exec Line: 4062796 TG reached the high limit. TG has been set to 190. Check patient positioning and if desired try Auto Prescan again or use Manual Prescan.

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 M5 Excite HDx HDi Platforms only.
Component
2208999-7 3.0T RF Body Coil and Case
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.