X100, MIS 28258A UPS Interface box to PDB TB1 Mislabeled for a X100 system

Symptom

X100, MIS 28258A UPS Interface box to PDB TB1 Mislabeled for a X100 system MIS 28258A UPS Interface box to PDB TB1 Mislabeled. Current MIS Chart in ASM (5160408-8 or 9-2EN rev1)shows PDB TB1 Wiring to be TB1 36, 37,38,39. The cable is labeled incorrectly and the Job card IST0065 section 4.3 shows the same incorrect info. The Correct wiring can be found on a older ASM 5160408-6-200. This ASM shows 39(Blk), 40(Grn), 41(Wht), 42(Red). MIS 28258A UPS Interface box to PDB TB1 Mislabeled. Current MIS Chart in ASM (5160408-9-2EN rev1)shows PDB TB1 Wiring to be TB1 36, 37,38,39. The cable is labeled incorrectly and the Job card IST0065 section 4.3 shows the same incorrect info. The Correct wiring can be found on a older ASM 5160408-6-200. This ASM shows 39(Blk), 40(Grn), 41(Wht), 42(Red).

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Component
SG-20 KVA UPS SG-20 KVA UPS Seelong. SG-20 KVA UPS OPM_SGS_USM_10K_40K_0US_V010.pdf 5380482 UL PDB Phase Control Relay LS1