GE Healthcare · MRExciteSigna2314498
HostID changed due to HP9300.
Symptom
The NVRAM HostID has Changed Value from the Previous Day Software will boot normally but no scan options are available. Error 2314498 will be recorded into the system error log everytime a scan is downloaded which the protocol requires scan options like FSE, GRE, etc. The hostID has changed from the original value from the day before.
Applies to
- Models
- Excite, Signa
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Component
- 5143798-2 HP 9300 Linux PC In FEMC this is "MR HDMR-2 Host". 5143798-2 HP 9300 Linux PC In FEMC this is "MR HDMR-2 Host".
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.