GE Healthcare · MRExciteSigna22481082244574
Unable to scan with DRF Over Range Errors.
Symptom
Unable to scan with DRF Over Range Errors APS received data over range errors from DRF2 on all channels (0-7), error number 2248108. Shortly after, there may or may not be an error indicating the master gradient failed to go to ready when commanded.
Error message
2248108 APS received data over range error from DRF, on chan= n. 2248108 TOiMR2 NSP APS:dataMoverTask 604 Exception Class: Idle Severity: Unknown Function: Acquisition/Pulse Generation: Acquisition PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 APS received data over range error from DRF, on chan = 2 Need to reduce receiver gain. 2244574 The GP reported a Gradient Amplifier % Axis Internal Wiring Fault or Control Undervoltage
Applies to
- Models
- Excite, Signa
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2396050-2 HFA CONTROL BOARD ASSEMBLY
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.