420V ACGD Breaker Trips Due To a Bad SGA Power Supply.

Symptom

Will Not Scan with Gradient Errors.

Error message

2244587 (breaker tripped) The GP is reporting SGA Power Supply Power Off. Check 420V power breaker on PDU. If breaker is tripped, replace SGA-PS. Check that 420V power line is fully seated between PDU & SGA-PS. Push EMO Reset Switch on PDU. Else, replace SGA-PS.

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Component
SecondOpinion ACGD SGA Power Supply 2334862-2 This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution