GE Healthcare · MRExcite31357
Unable to Send To PACS Due To Issue on PACS Problem External To GE MR System.
Symptom
Unable To Network To PACS
Error message
Remote Host XXXXXX is Down Retry after next retry_interval Error message from /export/home/sdc/logfiles/dcslog (the DICOM Server log): Thu Apr 10 08:02:45 2008 31357 dcspush.c:178 Remote Host GEPACS is Down - Retry after next retry_interval Thu Apr 10 08:05:55 2008 3938 dcspush.c:178 Remote Host GEPACS is Down - Retry after next retry_interval ... etc Note that "GEPACS" will vary depending on PACS remote hostname, and the first number after the date/time stamp is NOT an error code, rather is a PID process number, which will change every time.
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.