GE Healthcare · MRExcite22196632249267500212258106

Unable to run Auto Grafidy.

Symptom

Unable to Run Auto Grafidy Both Auto Grafidy and manual Grafidy would not run because of incorrect coil ID. The Grafidy coils are being recognized as SST. SST coils are not used in Profile.

Error message

2219663 Failed to write to the protocol file /usr/g/protocols/gep/other/.pn/grafidy_1. Wed May 14 11:03:52 2008 2219663 4 MR01 scn SrxProtUtil.c 1515Failed to write to the protocol file /usr/g/protocols/gep/other/.pn/grafidy_1.EN 177SR 1261210763034 0 1

2249267 Physiological Waveform Data Sequence Error. Function: Software/Firmware/Config: N/A PSD: XXX Coil: XX See Long Description Wed May 14 11:03:54 2008 2249267 MR01 NSPSCP:PhysDriver 631Exception Class: Unknown Severity: UnknownFunction: Software/Firmware/Config: N/APSD: grafidy Coil: SST Scan: 50021/1 Physiological Waveform Data Sequence Error. SCP-SPU link may have a problem. TPS Reset may be necessary. Correct Sequence number: 0x009BReceived sequence number: 0x009A->

2258106 Host coil validation error. Coil ID not found in coilconfig.cfg

Applies to

Models
Excite
Product
Profile with Excite
Software
ProfileHD Applies to Profile (Excite), MGD contains 1 VRE/ICN. Software is 14.x based.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.