GE Healthcare · MRExcite2225553
Intermittently Not Recognizing Coils and Multi coil TR Driver Faults Due to LPCA port A pins damaged.
Symptom
Intermittently Not Recognizing Coils and Multi coil TR Driver Faults Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225553 File: PMH.m 1.29 Method: spi_tr_mon Line: 6250 Multi-Coil T/R Driver Board Fault
Error message
2225553 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 3. Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225553 File: PMH.m 1.29 Method: spi_tr_mon Line: 6250 Multi-Coil T/R Driver Board Fault An open circuit was detected on port 3. Check cable to multi-coil board or check for an open pin diode.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 5111020-50 HDMR2 A plus Bezel and cable harness.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.