Unable to Perform UPM Functional Diagnostic due to UPM Reflected Power Cable from the Body Hybrid Not Seated At Rear Ped…
Unable to Perform UPM Functional Diagnostic due to UPM Reflected Power Cable from the Body Hybrid Not Seated At Rear Pedestal Cable Interface J4.
Symptom
Unable to perform UPM functional diagnostic, It reports UPM is not in proper mode. See Long Description Unable to perform UPM functional diagnostic, it reports "UPM is not in proper mode". There are no other errors posted or logged anywhere.
Error message
UPM is not in proper mode. UPM Functional diagnostic fails to run test: It reports "UPM is not in proper mode".
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.