GE Healthcare · MRSignaOvationExcite2225040
CWA ID103869. Unable to Download or Reset The TPS MGD or ISE
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2225040 IPG Communication Error. SPU failed to inform HPC that it had initialized correctly. Tue Dec 19 06:46:40 2000 Error: 2225040 Host: IRIS Process: NSP File: HPC_PORT_ID:hpc_init Line: 50 IPG Communication Error. Please reset TPS to use gating. SPU failed to inform HPC that it had initialized correctly. This could be caused by a bad SPU or bad DPR between SPU and the IPG main CPU.
Applies to
- Models
- Signa, Ovation, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.