Pular para o conteúdo principal
GE Healthcare · MR

Magnet Helium Vent kit shipped with LCC Magnets is Too Short.

Symptom

Magnet Helium Vent kit shipped with LCC and LCC300 Magnets is Too Short. The Standard Magnet Helium Vent kit shipped with LCC magnets is a 46-318057G6 and includes a 12 inch Vent Pipe.

Applies to

Product
Magnet LCC. Software Version Concept should NOT be selected for Magnet Products Serial #: 1.5T (Rxxx) / 1.0T (Qxxx) He Vessel Pressure:>3 psig but <5 psig Shield Temp:< 55 K Coldhead Temp:> 3.8K but < 4.5K Recondenser Temp:> 4.2K but < 4.55K Delta Temp (Coldhead - Recondensor):> 0 but < 0.3K Min. LHe Fill Level: 50% Min. LHe Ramp Level: 90% Min. LHe Shim Level: 75% Helium Volume:1940 L (1.5T) Helium Volume:1970 L (1.0T) Burst Disk <":2319285 Burst Disk >":2350415 Burst Disk O-ring:46-281101P1 Baffle Assy 2180818 Low Cryogen Consumption
Component
46-318057G1 VENT KIT ASSEMBLY 48 inch long Magnet Helium Vent Kit. See Long Description for FEMC Comments. FEMC Part Comments: Part #: 46-318057G1 Part Type: GP CONSUMABLES Part Weight: 12.73 LBR Repairable: No Returnable: No Hazardous Material: N Description: VENT KIT ASSEMBLY FRU: Yes Status: Orderable Parts Offered as: Current: 46-318057G1 Harvested: Not Available Exchange: Not Available Comments: FLS - HQ: 5 LOST IN-TRANSIT TO 0228FROM 0211 FEB. 1, 2001
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues