GE Healthcare · MRExcite222384922521062249418224931710819349562249307
SRF/TRF Board was hanging the MGD bus.
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2223849 APS EPROM power-up boot failed. (See attachment for extended error message) 2223849 lx-mr NSP TardisReset.cc 1807 APS EPROM power-up boot failed. 2252106 Failed to determine hardware state for APS board within 20 seconds. Error: 2252106 Host: BARLOW_MR Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for APS board within 20 seconds The board may be busy loading vxWorks or may be hung. Also check the serial cable for the board and the serial terminal server. 2249418 Network problem detected. Check MGD subnet network configuration. See Long Description Wed Apr 14 09:27:10 2004 2249418 2 CARD diagCoord mgdArp.c 157 Task ID: 0xf99df8 Time: 981 Object: SCP Network problem detected. Check MGD subnet network configuration. etherAddrResolve failed for network interface motfec0, IP 10.47.1.2, numTries 5, delay 100 ticks. status = -1. 2223849 SCP EPROM power-up boot failed (See attachment for extended error message) 2223849 rmny NSP TardisReset.cc 1841 Exception Class: Unknown Function: Host: NSP SCP EPROM power-up boot failed. Communication Timed-out from SCP board. Verify that MCP820 boards are installed and seated properly
Applies to
- Models
- Excite
- Product
- Profile with Excite
- Software
- Profile5 Applies to Profile5 (Excite) only. Profile5 is 12.x based software, not ProfileHD.
- Component
- 2280951 SRF/TRF board
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.