GE Healthcare · X-ray VascularAdvantxInnova3005030040

The Advatnx Console displays "Touch Reset to Continue".

Symptom

ADX: The Advatnx Console displays "Touch Reset to Continue". The Operator must touch the reset button to clear the error. The error 600E may be displayed on subsequent resets, or reboots. The Advantx Console displays "Touch Reset to Continue". The Operator must touch the reset button to clear the error. The error 600E may be displayed on subsequent resets, or reboots. This error occurs when overcurrent is sensed by the power unit (mA overload occurred), and there was less than 10 minutes since the previous mA overload. (If last mA overload happened more than 10 minutes before this one, error 42D9 is generated.)

Error message

ADX 30050 POWER UNIT : THYRISTOR TQ FAILURE 42DA 30040 OVERCURRENT SENSED. TOUCH RESET TO CONTINUE. AR ECM 0 0 HWStatus=0000 User=0000 0000 42DA XRTitan.cpp 810

Applies to

Models
Advantx, Innova
Product
Advantx E Monoplane Advantx E Systems advantx e, vamp Advantx E generator Systems
Software
All AdvantxE Software releases All AdvantxE Software releases All Advantx E Software releases
Area
X-ray Vascular