SecondOpinion ACGD SGA Power Supply 2334862-2.
Symptom
SecondOpinion ACGD SGA Power Supply 2334862-2 This part has been identified as one requiring second opinion approval before ordering. This part is frequently declared as FOA (Failed on Arrival), though the repair facility reports a high percentage of FOA’s are NDF (No Defect Found). Be sure to effectively isolate this part before ordering. See the resolution concept for the SecondOpinion for this part for steps to perform before ordering this part. This part has been identified as one requiring second opinion approval before ordering. This part is frequently declared as FOA (Failed on Arrival), though the repair facility reports a high percentage of FOA’s are NDF (No Defect Found). Be sure to effectively isolate this part before ordering. See the resolution concept for the SecondOpinion for this part for steps to perform before ordering this part.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion ACGD SGA Power Supply 2334862-2 This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.