Pular para o conteúdo principal
GE Healthcare · MRExcite2000023721205186740200581251000000001205186741760581251

Fails to Boot Applications Software and TPS reset Fails Due to Corrupt Software. 3T Excite G3. 200002372 The process ifc…

Fails to Boot Applications Software and TPS reset Fails Due to Corrupt Software. 3T Excite G3. 200002372 The process ifcc took too long to start up.

Symptom

System fails to boot Applications software and TPS reset fails. See long description. The system is a 1.5T LX, running 9.1 M4 software. The system will not boot applications. Looks like the Bit3 is down. Errors in the log look like TPS reset failure. Ping to tps fails as seen below: lx-mr 6# ping lx-mr-tps PING lx-mr-tps (216.33.27.2): 56 data bytes ping: sendto: Connection refused

Error message

200002372 The process ifcc took too long to start up. Fri Jun 10 08:09:19 2005 Error: 200002372 Host: lx-mr Process: hostMonitor File: ssProcStart.c Line: 904 The process ifcc took too long to start up.

Host watchdog timer expired, Aborting any current scan APS window during TPS Reset in mgd_term. Host watchdog timer expired. Aborting any current scan. CM_SO_TX_DAEMON: connec => 1205186740.200581251 : MSG_HANDLR2 NSP REQ : 011f 0009 00000000 : 006c - 0021 ffd3 003d 0004 0000 000a 7f7f 7f7f CM_SO_TX_DAEMON: connect to <ac14945a:05d4> failed, status = 22 The following packet is now lost: +------------------------------------------------------------+ | Revision = 2 Length (in words) = 108 | | Req. NMRID = MSG_HANDLR2 Opcode Number = 287 | | Rsp. NMRID = NSP Packet type = REQ | | Seq. # = 9 Status = 0x0 | | Endian = BIG | +------------------------------------------------------------+ 03a0dcf0: 0021 ffd3 003d 0004 0000 000a 7f7f 7f7f *.!...=..........* 03a0dd00: 7f7f 7f7f 7f00 0000 0000 0000 0000 0000 *................* 03a0dd10: 0000 0000 0000 0000 *................* 03a0dd20: 0000 0000 0000 0000 *................* 03a0dd30: 0000 0000 0000 0000 *................* 03a0dd40: 0000 0000 0

Applies to

Models
Excite
Product
3T Excite G3 Software ONLY
Software
G3 For 3T Excite only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.