GE Healthcare · MRExcite2238938
RF POWER AMP. Scan Stops due to Intermitent RF Amp fault 63.
Symptom
Scan Stops due to Intermittent RF Amp fault 63
Error message
2238938 3T RF Amplifier Non-volatile RAM Failure (fault 63). See long description. Wed Apr 23 12:46:31 2008 2238938 2 ami3t NSP SCP:../RfNbMdsHWO.cpp 2230 3T RF Amplifier Non-volatile RAM Failure (fault 63). Possible Causes: contents were corrupted or lost. Re-initialize RF Amplifier by turning 'Mains' breaker OFF, wait ten seconds, and turn back ON. If the fault persists, replace the RF amplifier.
Applies to
- Models
- Excite
- Product
- 3T Excite G3 Software ONLY
- Software
- G3 For 3T Excite only
- Component
- 2278810-4 3T 25KW RF AMPLIFIER KIT MKS ARX-X433 rev D 01480017ZZ RF POWER AMP model: 50-S26A-128 part#: 2278810-4
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required